Specification of Horiba OCMA-500 Oil Content Analyzer
Horiba OCMA-500 Oil Content Analyzer
OCMA-500 to measure concentration of oil contained in drainage and environmental water.
After injecting the water sample and solvent, all you have to do is press the start button,and the system will automatically conduct the monitoring operation from oil extraction to sample measurement and draining. With no more troublesome operations like opening/closing the drainage valve, monitoring is speeded up.In addition, the color graphic LCD and the backlit extraction tank have improved operability.
Specification
Model OCMA-500
Product name Oil content analyzer
Measurement method Solvent extraction - non-dispersive infrared absorption analysis method
Measured objects Substances extracted from sample water into solvent and having infrared absorption near a wavelength from 3.4 μm to 3.5 μm
Measurement range 0 mg/L to 200 mg/L
Resolution For mg/L 0 to 99.9: 0.1, 100 to 200: 1
For mg/g, mg/kg, mg/PC 0 to 9.99: 0.1, 10.0 to 99.9: 0.1, 100 to 200: 1
Repeatability 0 mg/L to 9.9 mg/L: ±0.3 mg/L.
10.0 mg/L to 99.9 mg/L: ±2.1 mg/L.
100 mg/L to 200 mg/L: ±5 mg/L. * For standard liquids
Display method 3.5 inches, 320×240 dots, Backlight Color graphic LCD
Calibration method Zero, span calibration
Amount of test sample required Sample water : Solvent = 2:1
Extraction solvent S-316
Amount of extraction solvent required 8 mL
Extraction method Built-in extractor
Ambient operating temperature 0°C to 40°C (no condensation)
Power supply AC 100 V to 240 V ±10%, 50/60 Hz
Power consumption AC 100 V: Approx. 60 VA, AC 240 V: Approx. 90 VA
External dimensions 342 (H) × 200 (W) × 313 (D) mm
Mass Approx. 7 kg
External output Output to a USB memory stick
Functions
300-item data memory
Self error determination
Stabilized measurement value display
Clock
Compliant with ASTM D7066-4