Test Method

The melting point of a crystalline solid is the temperature at which the solid to liquid phase transition occurs, referenced at one atmosphere (1 ATM) of pressure.

Specifications

Conforms to the specifications of:  

BP Appendix 5-Method 6; GLP 

Visual Image: 10x magnified displayed on monitor

Temperature Range: ambient + 5 to 315°C

Heating Rates: 0.2, 0.5, 1.0, 2.0, 3.0, 6.0, 12.0°C/min

Temperature Readability: 0.1°C

Cooling Time: 20 minutes (300°C to ambient)

Temperature Accuracy:

 

±0.5°C (ambient + 5 to 200°C)
±0.8°C (200 to 315°C)

Sample size: 5 mg (approximately)

Sample Holder:

  • Glass Slide ≤1mm ±0.02mm thick

Sample Cover: Glass Coverslip ±0.17mm thick

Temperature Sensor: Pt-100 (2 wire RTD)

Test Storage: Up to 20 tests with parameters

Untuk keterangan lebih lanjut, silahkan download PDF ini :