Specification of Koehler K87180 Semi-Automatic Vacuum Distillation System
Description
The Semi-Automatic Vacuum Distillation System features standard ASTM D1160 glassware enhanced with microprocessor control. The vacuum level, bath temperature and heating rates are programmable with up to 50 stored programs. Vapor temperature, distilling flask temperature and vacuum level are digitally displayed. Optional PC interface allows the distillation to be controlled from a PC and for data to be stored on the PC.
Test Method
Determines the range of boiling points for petroleum products that can be partially or completely vaporized at a maximum liquid temperature of 400°C at reduced pressures. The sample is distilled at a controlled, reduced pressure under conditions that are designed to provide approximately one theoretical plate fractionation. Initial and final boiling point is measured and a distillation curve relating volume percent distilled and the atmospheric equivalent boiling point temperature can be prepared.
Features
- Automatic Vacuum Control
- Automatic Heat Control
- Vapor Temperature Display
- Pot Temperature Display
- Automatic shutdown for high pot or vapor temperature
Specification
Conforms to the Specifications of:
ASTM D1160; ISO 6616
Distillation Temperature Range: Ambient to 400°C (752°F)
Condenser Temperature Range: Ambient +5°C to 150°C
Vacuum Range: 1.00 mmHg to 50 mmHg (0.13 to 6.7 kPa)
Electrical Requirements
220-240V 50/60Hz